已注冊用戶需要常規(guī)用戶名密碼登錄并綁定微信,才可以掃碼
點擊右上角可切換登錄模式
還沒有帳號?
現(xiàn)在注冊并享受以下權(quán)益:
- 直達官方商城在線購買
- 請求報價
- 參與精彩問答
- 獲取會員積分
- 兌換禮品
- 和更多 ...
CA610 Flouride Electrode Fill Procedure
CA610,在線氟化物
The fluoride ISE is the top electrode.
1.Unscrew the metal shell, remove the electrode and disconnect from the BNC connector.
2.Remove the tan colored O-ring from the flow cell.
3.Carefully wipe off the surface of the flow cell. Be careful not to lose the O-ring.
4.Slide the metal shell down the cable.
5.Unscrew the tip of the electrode from the body of the electrode.
6.Upon removal of the electrode tip, inspect the sealing O-ring at the base of the threads on the body. Replace if torn or damaged . It is recommended that the O-ring (Product # 1325300) be replaced with every new tip.
7.Fill a syringe with fluoride electrode fill solution (Product # 4450126).
8.Set the tip of the electrode on a table, face down.
9.Fill the tip full of fill solution.
10.Hold the electrode body, open end up.
11.Insert the syringe tip down into the electrode (be careful not to damage the wire in the electrode).
12.Fill the electrode body to the top, then continue injecting solution as you remove the syringe tip from the electrode.
13.Turn the electrode body upside down - open end down - and screw it into the electrode tip.
14.Blot off any drips on the electrode.
15.Do not wipe totally dry! Excessive wiping can create a static charge on the electrode that will cause erroneous readings!
Hach has a complete portfolio of instruments and chemistries with support and services to help you get the right results.
了解更多
隱私政策 Cookie通知 / 滬ICP備13034148號-4 / 滬公網(wǎng)安備31010502004971號 / 滬(浦)應(yīng)急管危經(jīng)許[2023]201871
? 版權(quán) 2015-2024 哈希中國版權(quán)所有 / 哈希水質(zhì)分析儀器(上海)有限公司 / 本網(wǎng)站用字經(jīng)北大方正電子有限公司授權(quán)許可